Abnormal status indicator
 Abnormal status indicator
SOFTWARE

Abnormal status indicator

Regarding the accessory issue, we suggest purchasing the required components and then, under the guidance of the after-sales department, you can replace them yourself. Please note that the specific steps for replacement may vary depending on the different devices.
Channel protection prompts serial port error
Channel protection prompts serial port error
SOFTWARE

Channel protection prompts serial port error

Solution 1: Check whether the network cable/ethernet port connection is poor.
Solution 2: Power off for 20 minutes, then restart the device and computer.
Channels not displayed after power failure
Channels not displayed after power failure
SOFTWARE

Channels not displayed after power failure

Solution 1: End the channel status before power-off, and then restart the channel.
Solution 2: It may be caused by database damage, and a script is needed to repair the database. Please contact the after-sales department for database repair.
Channel connection failed
Channel connection failed
SOFTWARE

Channel connection failed

Solution 1. Shut down the equipment for 20 minutes, then restart both the equipment and the computer.
Solution 2. Upgrade the client-server, and format the control unit (upgrading the server and formatting the control unit may pose risks, it is recommended to contact the after-sales department for assistance).
Solution 3. It may be due to the channel stop time being too long, resulting in a channel continuation alarm. The control unit time synchronization has not been updated completely. After synchronization is completed, the channel continuation will return to normal.
Channels automatically stop
Channels automatically stop
SOFTWARE

Channels automatically stop

Clean up disk space or replace the SD card.
Channels cannot be started/stopped
 Channels cannot be started/stopped
SOFTWARE

Channels cannot be started/stopped

First, check the physical connections of the equipment (network cable, fixture, flight plug and etc.).
If the inability to start/stop is caused by communication issues, then restart the equipment.
Mapping error, startup failure
Mapping error, startup failure
SOFTWARE

Mapping error, startup failure

Solution 1: Check if the network cable and port are well connected. Solution 2: Restart the host computer, change the boot sequence (start the lower computer before the host computer), and then remap.
Remapping correction plan: If the mapping is incorrect while the channel is in the running state and cannot be reset, first pause the channel, save the data, then right-click and select "Clear Flag," enter "Mapping Settings," right-click and select "Undo Current Channel," then remap the channel and save the mapping.
Solution 3: Modify the equipment box number through calibration software/box number modification software (please contact after-sales support to resolve the issue).
Large voltage & current fluctuations
Large voltage & current fluctuations
SOFTWARE

Large voltage & current fluctuations

Solution 1: Check if the fixture has poor contact, readjust or replace the fixture as needed.
Solution 2: Re-calibrate the equipment. For calibration, please contact the after-sales department to resolve.
Charging and discharging time is incorrect
Charging and discharging time is incorrect
SOFTWARE

Charging and discharging time is incorrect

Solution 1: Check if the set voltage and current match the displayed voltage and current.
Solution 2: Verify the accuracy of the equipment. (If there is a discrepancy, calibration service may be required. Please contact the after-sales department for resolution).
Discharge cannot reach full capacity
Discharge cannot reach full capacity
SOFTWARE

Discharge cannot reach full capacity

Check if the negative power supply is working properly. Devices without a negative power supply will have lower current as the voltage decreases (devices with 6A or above have a negative power supply). Verify if the displayed and set voltage and current values are consistent.
Abnormal high-rate discharge capacity
Abnormal high-rate discharge capacity
SOFTWARE

Abnormal high-rate discharge capacity

Solution 1. Check if the fixture has poor contact.
Solution 2. If the equipment tests battery capacity decay rapidly, check the equipment accuracy and test chamber performance.
Solution 3. Upgrade the control unit program. Upgrading the control unit may have risks, please contact the after-sales department for assistance.
Slow charging response time
Slow charging response time
SOFTWARE

Slow charging response time

It may be caused by a software program error. Please contact the after-sales department for assistance.
Charging does not switch/Process step cannot switch/Abnormal switching
Charging does not switch/Process step cannot switch/Abnormal switching
SOFTWARE

Charging does not switch/Process step cannot switch/Abnormal switching

Solution 1:Check if the process step settings are correct. Then inspect the fixture for any damage.
Solution 2:Verify if the displayed current and voltage values match the set values. If not, please contact the after-sales department.
The channel indicator light status is inconsistent with the test status
The channel indicator light status is inconsistent with the test status
SOFTWARE

The channel indicator light status is inconsistent with the test status

It may be necessary to replace the accessories. We recommend contacting the after-sales department to purchase the necessary accessories and replace them yourself.
Unable to perform cycle testing
Unable to perform cycle testing
SOFTWARE

Unable to perform cycle testing

Check if the process step settings are correct. If it is confirmed that the inability to test is not caused by process step setting issues, please contact the after-sales department for assistance.
Battery swelling during testing
Battery swelling during testing
SOFTWARE

Battery swelling during testing

Check the data and verify if the displayed current and voltage values match the set values.