CT-4008Tn-5V12A-S1 Manual Current and Voltage Calibration Guide

Latest updated: Nov 12, 2026

BTS manual current and voltage calibration guide

1. Manually calibrate current

1.1 Connecting the power cable

First, connect the power cables of the multimeter, Control Unit, and NEWARE Tester.

Power cable connection for multimeter, Control Unit and Tester

1.2 Connecting the network cable

Connect the Control Unit to the Tester using a network cable. The Control Unit is connected to the switch. The multimeter is connected to the switch.

Finally, connect the switch to the computer using a network cable.

Network wiring diagram for manual calibration

1.3 Connecting the multimeter and the Tester

Next, connect the multimeter test line, red represents the positive pole and black represents the negative pole.

The red wire is connected to the positive terminal. The black wire is connected to the negative terminal. Connect the fixture to the Tester and tighten it.

Connect the fixture to the test wire on the multimeter.

Multimeter test lead and fixture wiring

1.4 Start the device and check the multimeter IP

After powering on the device, start the Tester, Control Unit, and multimeter in turn.

Then click Shift key, Display key, I/O key, LAN Settings key. You can view the multimeter IP address.

Check multimeter IP via LAN Settings

Press Shift and DCV to switch to DCI mode. The multimeter screen displays the current unit.

Switch multimeter to DCI DC current mode

1.5 General setup section of the computer

1.5.1 Turning off protection and firewalls

Enter Windows Security. Find "Manage settings" under "Virus & threat Protection".

Turn off "Real-time protection", "Cloud-delivered protection", and "Automatic sample submission".

Disable Windows real-time protection

Click on "Firewall & network protection" in the left column. Turn off "Domain network", "Private network", and "Public network".

Turn off Windows firewall for all networks

1.5.2 Setting the computer IP

First, access the "Control Panel". Click on "Network and Internet" and then proceed to the "Network and Sharing Centre". In the left column, click on "Change adapter settings".

Right-click on "Ethernet", select "Properties", and choose the "Internet Protocol Version 4 (TCP/IPv4)" item.

Open Ethernet IPv4 properties

Click on "Properties" and select "Use the following IP address". Manually enter the IP address as "192.168.10.250". It is important to note that the IP address must match the IP address of the Control Unit server.

Click on the input box for "Subnet mask" to automatically detect the mask address.

Finally, click "OK" and then click "Close".

Set static IPv4 address 192.168.10.250

1.6 Guidance on BTS software operation

1.6.1 User login and channel mapping

Open NEWARE BTS software, click "User", select "User log in". The user name is "admin" and the password is "neware".

NEWARE BTS admin login page

Right-click in the blank area and select "Reset mapping". Right-click and select "Main CH default mapping 1x8". Then right-click to save the mapping.

Reset channel mapping menu

Save channel mapping

The software interface displays 8 test channels.

8 test channels on BTS main interface

Click on "Tools" and select "Calibration". After clicking "Connect", the channel and multimeter IP will be displayed in the left column. Ensure that the multimeter IP address is the same as that on the multimeter display.

If the IP address of the multimeter is not displayed, click "Connect" again.

Calibration tool connect multimeter by IP

Click "System configuration" and select "Standard connection settings". You can view and modify the multimeter's IP address here. Confirm that the following two options are unchecked, then click "OK".

Multimeter standard connection settings

1.6.2 Calibration CC (mA)

Select "Manual-Adjust" and then select the channel that you want to calibrate (shown here as channel 1). This is the calibration current, so double click "CC(mA)".

After successful selection, a yellow arrow is displayed before the option.

Select CC(mA) for manual current calibration

And then click Start. Enter the multimeter reading at this time into the "Multimeter Value" box, noting the conversion unit. Then click "Fill in" next to "Display Value" in the "Low Point" box below.

Fill low point value for CC calibration

Then click Start. Enter the current multimeter reading into the "Multimeter Value" box. Click "Fill in" next to "Display Value" in the "High Point" box below.

Fill high point value for CC calibration

Do this in turn and complete all current calibration items.

1.6.3 Calibration DC(mA)

When the device is not equipped with a negative power supply, an external power supply is required for calibrating "DC(mA)".

If the device serial number includes the letter 'F', it indicates that the device comes with a negative power supply; otherwise, it does not.

The hardware connection mode is shown in the figure, and the software operation is the same as calibrating "CC(mA)".

External power supply wiring for DC(mA) calibration

2. Manually calibrate voltage

2.1 Replacing wiring

Replace the red wire on the multimeter into the Input jack.

Multimeter red wire to Input jack

Then, connect an external cement resistor. Attach the crocodile clamps to the two ends of the cement resistor and the positive and negative probe terminals of the multimeter respectively.

Cement resistor wiring setup

Multimeter probe connection for voltage calibration

Next, switch from DCI mode to DCV mode, and the units on the screen will change to voltage.

Switch multimeter to DCV voltage mode

2.2 BTS software operation

2.2.1 Calibration CV (V)

Also select "Manual-Adjust" and then select the channel that you want to calibrate (shown here as channel 1). This calibration is for voltage adjustment, so double-click the "CV(V)" option.

Upon successful selection, a yellow arrow will be displayed next to the option.

Select CV(V) for manual voltage calibration

Now, click "Start". Enter the current multimeter display value in the "Multimeter Value" box, noting the conversion unit. Then click "Fill in" next to "Display Value" in the "Low Point" box below.

Fill low point value for CV calibration

Afterward, click "Start". Enter the current multimeter reading into the "Multimeter Value" box again. Click "Fill in" next to "Display Value" in the "High Point" box below.

Fill high point value for CV calibration

Repeat these steps for all the voltage calibration items.

2.2.2 Calibration DV(V)

To calibrate DV, you need to connect a constant voltage power supply in series, as shown in the figure. The software operation is the same as calibrating "CV(V)".

Series constant voltage power supply wiring for DV(V) calibration

3. Exported data

Select the completed items, and right-click to manually export the calibration data.

Export manual calibration report


neware-battery-test-newareAI neware-battery-test-newareStore neware-battery-test-neware-newell

Find the Right Battery Test Equipment for Your Needs.

Application Scenarios

Trusted testing solutions for global clients.

Solid-State Battery Research - NEWARE Solid-State Battery test
Solid-State Battery Research

The lab focuses on solid-state battery research to overcome traditional lithium batteries' safety and energy density issues, supporting environmental sustainability. It develops innovative solid-state electrolytes, refines electrode materials, and investigates ion transfer and interface stability to revolutionize battery technology.

View more
Electric Vehicle Battery
Electric Vehicle Battery

The electric vehicle battery industry is rapidly developing, focusing on technological innovation, market competition, and sustainability. Research hotspots include solid-state batteries, new types of electrolytes, BMS optimization, and recycling technologies. The environmental adaptability, safety, and economic viability of batteries are key research areas, and the industry is expected to undergo more innovation and transformation.

View more
Battery Materials Research - NEWARE battery test
Battery Materials Research

We specialize in battery preparation technology research, focusing on overcoming existing energy storage challenges by innovating in electrode materials, battery chemistry, and manufacturing processes to improve performance, enhance safety, and reduce costs. Sustainability and recycling technologies for batteries are also emphasized to mitigate environmental impacts and foster the growth of green energy.

View more
Energy Storage Battery Testing Solution
Energy Storage System

To power the energy transition, the storage industry is evolving towards large-scale, high-quality development, focusing on safety, efficiency, and lifecycle value over mere price competition.

View more
NEWARE and Cookies
We use cookies to personalize and improve your experience with our website. By continuing to browse the site you are agreeing to our use of cookies.