The channel in testing has transitioned to the completed state
The channel in testing has transitioned to the completed state
SOFTWARE

The channel in testing has transitioned to the completed state

Solution 1:Shut down the equipment for 20 minutes, then restart the device and computer.
Solution 2:Restart the control unit (restart on the device is sufficient).
Jump to complete as soon as the step test is started
Jump to complete as soon as the step test is started
SOFTWARE

Jump to complete as soon as the step test is started

It may be caused by incorrect initial step settings or poor battery contact.
Please check if the initial step settings are correct and ensure that the fixture has good contact with the battery.
Failed to execute the step file
Failed to execute the step file
SOFTWARE

Failed to execute the step file

Check the Ethernet cable and Ethernet port for good contact, shut down the device for 20 minutes, restart the device, and restart the computer.
Not following the set steps
Not following the set steps
SOFTWARE

Not following the set steps

The steps saved using a higher version of the software cannot be launched on an older version. It is recommended to update the software version. For example, process step files saved in BTS8.1 cannot be launched in BTS8.0.
Unable to reset the step
Unable to reset the step
SOFTWARE

Unable to reset the step

You can try restarting the computer first to resolve the issue. If the problem is not resolved, please upgrade the client-server. If you are unsure how to upgrade the client-server, please contact the after-sales department.
Steps cannot be saved
Steps cannot be saved
SOFTWARE

Steps cannot be saved

You can try restarting the computer first to resolve the issue. If the problem is not resolved, please upgrade the client-server. If you are unsure how to upgrade the client-server, please contact the after-sales department.
Step cannot be invoked
Step cannot be invoked
SOFTWARE

Step cannot be invoked

You can try restarting the computer first to resolve the issue. If the problem is not resolved, please upgrade the client-server. If you are unsure how to upgrade the client-server, please contact the after-sales department.
Prompt: No NDA data
Prompt: No NDA data
SOFTWARE

Prompt: No NDA data

Check the computer's running memory, it should not exceed 90%. Clean up the computer's disk space. Pause the test first, then save the data separately. Manually delete historical data, it is recommended to delete data older than six months. Add a new hard drive to the computer and retest.
Data cannot be saved
Data cannot be saved
SOFTWARE

Data cannot be saved

Solution 1: If the control unit program is abnormally stuck, upgrade the control unit program.
Solution 2: Delete unnecessary backup tasks, modify backup settings, clean up disk space, or replace (add) new hard drives.
Channels cannot be stopped on the software interface
Channels cannot be stopped on the software interface
SOFTWARE

Channels cannot be stopped on the software interface

Restart the control unit (restart the device), wait until the control unit screen is completely off, and then turn it on again.
Software lag/freeze
Software lag/freeze
SOFTWARE

Software lag/freeze

Check the computer's running memory, end processes occupying memory, and ensure that the memory usage rate does not exceed 90%. After disabling the firewall and antivirus software, restart the software. If the issue persists, then reinstall the software. Before reinstalling, please uninstall or disable the firewall and antivirus software.
Software interface displays garbled characters
Software interface displays garbled characters
SOFTWARE

Software interface displays garbled characters

After uninstalling the software, reinstall it. If the problem still persists, please contact the after-sales department for assistance.
Software font display abnormality
Software font display abnormality
SOFTWARE

Software font display abnormality

You can try upgrading the software to resolve the issue. If the problem persists, it may be a programming error.
The voltage or current is not at the set value
The voltage or current is not at the set value
DATAS

The voltage or current is not at the set value

Solution 1: The device does not come with a negative power supply and should not be discharged to 0V or discharged with constant current at low voltage. Check if the settings are incorrect.
Solution 2: Check if the device version and client version are the same.
Steps to view the client version: Open the BTS8.0 software, right-click and select "View Channel Information" → you can see "Server Version" and "Client Version".
No voltage, no current
No voltage, no current
DATAS

No voltage, no current

Check if the battery and fixture are in good contact and inspect whether the aviation plug wire is intact and properly connected to the equipment.
Voltage or current exceeds the upper limit
Voltage or current exceeds the upper limit
DATAS

Voltage or current exceeds the upper limit

Solution 1: Check the process settings (voltage and current upper and lower limits) to ensure they are reasonable.
Solution 2: The equipment needs to be recalibrated.